• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ENSITE PRECISION¿ POWER CABLE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ENSITE PRECISION¿ POWER CABLE; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100143451
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Event Description
Following a completed atrial fibrillation ablation procedure with no patient consequences, the inner metal prongs for the power cable of the precision link were noted to be melted.There were no adverse consequences to the user of the device.
 
Manufacturer Narrative
One ensite precision¿ power cable was received for analysis.Visual inspection of the returned power cable revealed physical damage to the collar housing on the redel and 4 pins were sheared off which would have led to the field reported event.It is unknown how the physical damage occurred.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, the reported event was confirmed.The cause for the reported event was isolated to physical damage to the connector housing of the redel connector and 4 pins were sheared off due to an unknown event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE PRECISION¿ POWER CABLE
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12888648
MDR Text Key281382365
Report Number2184149-2021-00383
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100143451
Device Lot Number6702673
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2021
Initial Date FDA Received11/29/2021
Supplement Dates Manufacturer Received01/05/2022
Supplement Dates FDA Received01/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-