Model Number TN,DXH 500
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2021
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported recovering high platelet (plt) values for patient samples post preventative maintenance (pm) when run on their dxh500 hematology instrument.There was no report of an adverse event.There was no report of erroneous patient results reported outside the laboratory.There was no impact to patient treatment.Patient data was requested but not provided for review.
 
Additional Manufacturer Narrative
Patient data was not provided by the customer.The field service engineer (fse) guided the customer to perform repeatability and values recovered were satisfactory.The fse also performed calibration and updated the instrument with the new calibration factors resolving the reported issue.Bec internal identifier - (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXH500 HEMATOLOGY INSTRUMENT
Common Device Name
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER IRELAND INC.
lismeehan
o¿callaghan¿s mills
co. clare
EI 
Manufacturer (Section G)
BIT GROUP FRANCE
rue de la valsière
parc euromédicine ii
montpellier cedex 34090
FR   34090
Manufacturer Contact
harry long
1000 lake hazeltine drive
m/s r590c
chaska, MN 55318
9523681224
MDR Report Key12888750
Report Number1061932-2021-00164
Device Sequence Number4335447
Product Code GKZ
UDI-Device Identifier15099590671242
UDI-Public(01)15099590671242(11)NO-DATA
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K181475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 11/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTN,DXH 500
Device Catalogue NumberB40601
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2021
Initial Report FDA Received Date11/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-