• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3013-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2021
Event Type  malfunction  
Event Description
It was reported that during a routine check performed by a getinge field service engineer (fse), the cs100 intra-aortic balloon pump (iabp) was showing low battery.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp unit for the reported battery issue and was able to reproduce the reported issue.To fix the issue the fse replaced the battery and verified that it was charging, he then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.(b)(6).The initial reporter named is a getinge employee who has different contact details from that of the event site.A contact person at the event site is (b)(6), biomed.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: d1, d4, g1 (contact person ¿ mfg site).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,DOM
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12890181
MDR Text Key283620509
Report Number2249723-2021-02749
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107745
UDI-Public10607567107745
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-UC-3013-53
Device Catalogue Number0998-UC-3013-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/08/2021
Initial Date FDA Received11/30/2021
Supplement Dates Manufacturer Received10/10/2022
Supplement Dates FDA Received10/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-