H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultraverse 014 pta dilatation catheter was returned for evaluation.No specific anomalies noted on the visual evaluation.On the functional testing of the returned device upon inflation using an in-house presto inflation device the balloon started leaking, upon microscopic observation it was noted a longitudinal balloon rupture.No further functional testing performed.Therefore, the investigation for the reported balloon rupture was confirmed as the balloon started leaking from the longitudinal rupture during the functional testing.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 12/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|