| Model Number |
BI70002000 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) 2021, (b)(4), (rep, for): medtronic received information regarding an imaging system being used outside of a procedure during an installation.It was reported that the imaging system would not boot into 2d image acquisition functionality and the motion bar scrolled continuously.Additionally it was reported that a movement error occurred and there was no 2d/3d imaging.Error logs found an error with the positioner, additionally, wiring was found to be tight to the positioner's and interface boards.A rj45 socket on the unit appeared to be cracked and the connection was intermittent from the positioner board, resulting in a boot failure.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
Onsite functional and visual examination was performed by a manufacturer representative.The positioner was replaced and re-ran cable loom and the issue was resolved.The system passed a system checkout and was returned to an operational condition.The representative also found a damaged cover.Concomitant medical products : other relevant device(s) are: kit svc o2 bi71000443 pos mot ctrl namp.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: bi71000443r, serial/lot : (b)(6) rev.2 h3: hardware analysis confirmed the reported issue.When the returned position motion controller was installed in a known good system, the motion did not ready.The system did not initialize and was unable to boot and ready.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|