• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. SAFE LOCK APD LUER LOCK CONNECTOR; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. SAFE LOCK APD LUER LOCK CONNECTOR; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Model Number 050-95018
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the apd luer lock connection during dwell 1 of 4 of their pd treatment.There were no alarms associated with the fluid leak.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to cancel treatment and re-setup the cycler with new supplies.Upon follow up, the patient stated they did not complete treatment on the night of the fluid leak.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient confirmed they are continuing peritoneal dialysis therapy with no further issues.The luer lock connector used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Additional Manufacturer Narrative
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the apd luer lock connection during dwell 1 of 4 of their pd treatment.There were no alarms associated with the fluid leak.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to cancel treatment and re-setup the cycler with new supplies.Upon follow up, the patient stated they did not complete treatment on the night of the fluid leak.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient confirmed they are continuing peritoneal dialysis therapy with no further issues.The luer lock connector used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Event or Problem Description
It was reported that a peritoneal dialysis (pd) patient discovered a fluid leak from the apd luer lock connection during dwell 1 of 4 of their pd treatment.There were no alarms associated with the fluid leak.It is unknown at which point in therapy the leak may have begun.The cause of the leak is unknown.The patient was advised to cancel treatment and re-setup the cycler with new supplies.Upon follow up, the patient stated they did not complete treatment on the night of the fluid leak.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient confirmed they are continuing peritoneal dialysis therapy with no further issues.The luer lock connector used by the patient was discarded and is not available for return for physical evaluation by the manufacturer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFE LOCK APD LUER LOCK CONNECTOR
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
director, quality systems
1100 e, military hwy, suite c
pharr TX 78577
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key12893966
Report Number8030665-2021-01789
Device Sequence Number11649567
Product Code KDJ
UDI-Device Identifier00840861100811
UDI-Public00840861100811
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K124024
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/30/2022
Device Model Number050-95018
Device Catalogue Number050-95018
Device Lot Number21HR08084
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2021
Supplement Date Received by Manufacturer12/09/2021
12/22/2021
Initial Report FDA Received Date11/30/2021
Supplement Report FDA Received Date12/21/2021
12/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELFLEX PD FLUID; DELFLEX PD FLUID; DELFLEX PD FLUID; LIBERTY CYCLER SET; LIBERTY CYCLER SET; LIBERTY CYCLER SET; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER; LIBERTY SELECT CYCLER
Patient Age67 YR
Patient SexMale
Patient Weight89 KG
-
-