• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FULL RADIUS BLADE 3.5MM DISP; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FULL RADIUS BLADE 3.5MM DISP; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 7205305
Medical Device Problem Code Flaked (1246)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, during an arthroscopy procedure, the full radius blade created metal flakes immediately after use.The procedure was completed with a backup device.It is unknown if there was a delay.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection revealed the device was not returned with original packaging and the device has debris on it.No physical damage is visible to the device.A functional evaluation of the returned device found that device was ran and no metal flakes were found coming off device.Device passed all functional test.A complaint history review concluded this was an isolated event.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.It was determined the device did not contribute to the reported event.The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL RADIUS BLADE 3.5MM DISP
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12894113
Report Number1219602-2021-02609
Device Sequence Number19697748
Product Code HAB
UDI-Device Identifier03596010251336
UDI-Public03596010251336
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2026
Device Model Number7205305
Device Catalogue Number7205305
Device Lot Number50925533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2021
Supplement Date Received by Manufacturer03/02/2022
Initial Report FDA Received Date11/30/2021
Supplement Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-