| Model Number |
66801069 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Skin Disorders (4543)
|
| Date of Event |
10/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy (acl or other), upon removal of the allevyn life l 15.4x15.4 ctn10, it was observed that the patient's skin was macerated.Also, according to client´s description there was a lot of adhesive residue.Treatment was resumed, with a greater 30 minutes delay, with a back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, used in treatment, was not returned for evaluation.All provided information has been reviewed and we have not been able to establish a relationship between the device and the reported event or determine a root cause.A probable root cause may include a component failure, the wound contact layer within these dressing can be affected by storage temperature fluctuations as detailed in the ifu.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.Since no patient injuries were reported, no further clinical/medical assessment is warranted.The instructions for use contain comprehensive instructions on the safe operation and use of the device.The risk files mitigate the reported issue with no updates required.A complaint history review has found other related events, with corrective action implemented related to the reported event.Smith + nephew will continue to monitor for adverse trends.
|
| |
|