| Model Number |
03P6832 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The account generated false elevated architect magnesium that repeated lower after recalibration.On (b)(6) 2021, sid (b)(6) generated architect magnesium of 3.1 mg/dl that was reported out of the lab.The account recalibrated and repeated the sample with architect magnesium of 1.3 mg/dl that was reported with a corrected report.The account uses a magnesium normal range of 1.8 to 2.6 mg/dl.No specific patient information was provided.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient identifier does not contain enough character spaces, the entire id is (b)(6).An evaluation is in process.A followup report will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided that expired architect magnesium reagent lot 42219un20 used on (b)(6) 2021 for testing patient samples.The architect magnesium package insert instructs the user not use reagents beyond the expiration date.Based on the additional information, this event is no longer reportable and no further followup will be provided.
|
| |
|
Event or Problem Description
|
|
Additional information was provided that expired architect magnesium reagent lot 42219un20 used on (b)(6) 2021 for testing patient samples.
|
| |
|