• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TIGHTROPE XP BUTTRESS PLATE IMPLANT SYS; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TIGHTROPE XP BUTTRESS PLATE IMPLANT SYS; WASHER, BOLT NUT Back to Search Results
Model Number TIGHTROPE XP BUTTRESS PLATE IMPLANT SYS
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a syndesmosis repair surgery with internal brace the device did not work properly.After tightening the sutures the other suture was cut and the surgeon could not tighten it.The device broke inside the patient and all broken pieces were retrieved from the patient.He needed to change the device to a single packed tightrope implant to continue with the procedure.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device (ar-8926t).It was not necessary to switch the surgical technique or do a second surgery.
 
Event or Problem Description
It was reported that during a syndesmosis repair surgery with internal brace the device did not work properly.After tightening the sutures the other suture was cut and the surgeon could not tighten it.The device broke inside the patient and all broken pieces were retrieved from the patient.He needed to change the device to a single packed tightrope implant to continue with the procedure.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device (ar-8926t).It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIGHTROPE XP BUTTRESS PLATE IMPLANT SYS
Common Device Name
WASHER, BOLT NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12896568
Report Number1220246-2021-04014
Device Sequence Number5365814
Product Code HTN
UDI-Device Identifier00888867352278
UDI-Public00888867352278
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K201522
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model NumberTIGHTROPE XP BUTTRESS PLATE IMPLANT SYS
Device Catalogue NumberAR-8959TDS
Device Lot Number13403360
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2021
Initial Report FDA Received Date11/30/2021
Date Device Manufactured05/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-