Model Number UNI GLENOID-PERIPHERAL LOCK SCRW
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a universe reverse shoulder surgery when trying to insert the 30 mm screw into the baseplate & bone the screwdriver screwed through, as the screw was too tight.The screwdriver was defective and the drivers became distorted.The screw head had twisted, could no longer be moved and could not be fully inserted.The surgeon knocked the screw flat and inserted the glenosphere.No part of the devices broke off.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with the same devices.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
Complaint not confirmed, the screw was not returned and no picture provided.The x-ray provided did not provide enough information to confirm or investigate the complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI GLENOID-PERIPHERAL LOCK SCRW
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12896739
Report Number1220246-2021-04016
Device Sequence Number20267376
Product Code PHX
UDI-Device Identifier00888867057722
UDI-Public00888867057722
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUNI GLENOID-PERIPHERAL LOCK SCRW
Device Catalogue NumberAR-9145-30
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer11/10/2021
Initial Report FDA Received Date11/30/2021
Supplement Report FDA Received Date12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-