• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. TENO SCRW,BIO-COMP; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. TENO SCRW,BIO-COMP; SCREW, FIXATION, BONE Back to Search Results
Model Number AR-1530BC
Health Effect - Clinical Code Pain (1994)
Date of Event 11/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a plantar plate repair with cpr kit and an edb transfer using the mini biotenodesis multiple issues occurred with several devices.The surgeon reported that cpr went well until fixing the weil¿s osteotomy.Initially the surgeon struggled to snap off the twist off screw.The first one he only partially bent it on hen engaged the drill so it span around.He then tried another screw in a different position and it seemed fine.After the first screw for the edb transfer went in find the second tenodesis screw did not.The surgeon kept pushing and twisting but no engagement beyond a certain point.Screwdriver snapped leaving a piece of wire embedded in the screw.He then tried to twist further but screw came out.He tried another one but after inserting the screwdriver handle would not disengage from the screw.The wire holding the tendon for transfer snapped and therefore this part of the procedure did not work.The surgeon tried another twist off but it did not snap.It was further reported that due to these failures the bone was damaged.This meant there wasn¿t space or quality of bone left to finish the procedure.17-nov-2021 update: further information were provided that the surgery started with the weil¿s osteotomy, then the plantar plate repair.This all went smoothly.Then the surgeon isolated the edb tendon and drilled 2 x 2.5mm holes using the mini biotenodesis disposables set ar-1530ds, he then inserted a 3x8mm biotenodesis screw (ar-1530bc) in the proximal hole which went fine.The surgeon then tried to insert the second screw in using an identical product but the screw would not engage with the tendon and hole.When he had inserted it and tried to disengage the screw driver, the screw came out with the screw driver.He tried this a couple more times until it finally worked.However it was slightly prominent, he went to tighten it and the screw came out again.He had by this point trimmed the tape and suture on the end of the tendon so was unable to tension again to finish that aspect of the procedure.The surgeon reported that the procedure could not be finished as planned and it is unclear if the plantar plate repair is enough to alleviate the patient pain.The following devices were used during the procedure: ar-1530ds-1 lot: 12161979.Ar-1530bc lot: 12649748.Ar-1530bc lot: 11793696.Ar-8930-12s lot: 10574389.Ar-8690ds lot: 10300450.Ar-8930-11s lot: 10350805.Ar-8941ks lot:11790344 x3.Ar-7232-03-01 lot: 19272.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENO SCRW,BIO-COMP
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key12896853
Report Number1220246-2021-04018
Device Sequence Number11937714
Product Code HWC
UDI-Device Identifier00888867018600
UDI-Public00888867018600
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K122374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberAR-1530BC
Device Catalogue NumberAR-1530BC
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2021
Initial Report FDA Received Date11/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-