Model Number FCS17
Medical Device Problem Code Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the cutting & coagulation was not working.The surgery was delayed by 15 minutes.The device was replaced.The procedure was completed successfully.
 
Additional Manufacturer Narrative
(b)(4).Batch #: r4029j.Additional information was requested and the following was obtained: did the patient experience any adverse consequences due to this issue?-no investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.The device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and can result in not passing the pre-run test followed by an alert screen, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure, and continued usage can result in a broken blade.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HS FOCUS 17 CM CURVED SHEAR
Common Device Name
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key12897023
Report Number3005075853-2021-07239
Device Sequence Number4024559
Product Code LFL
UDI-Device Identifier10705036001942
UDI-Public10705036001942
Combination Product (Y/N)N
PMA/510(K) Number
K063192
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberFCS17
Device Catalogue NumberFCS17
Device Lot NumberR4029J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/09/2021
Initial Report FDA Received Date11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/30/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR AND HANDPIECE
Patient SexUnknown
-
-