| Catalog Number |
11895354160 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The test strips were requested for investigation.The retention material of lot 51475500 was visually checked and qc documentation was checked.The retention material showed no abnormalities and fulfill the requirements.Qc documentation showed no abnormalities.
|
| |
|
Event or Problem Description
|
|
The initial reporter received a questionable leukocytes result for one patient urine sample with a urisys 1100 urine analyzer serial number (b)(4).The leukocyte result from the meter was negative.The patient sample was sent for a laboratory culture that reported >100,000 e.Coli.No urinalysis dipstick test was performed at the laboratory.The questionable result was reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
No product was returned for investigation.The investigation did not identify a product problem.The cause of the event could not be determined.
|
| |
|