Model Number UM-S20-17S
Device Problems Defective Device (2588); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/04/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was received and evaluated.Device evaluation confirmed the customer reported issue.Inspection found probe was damaged with crush and bubbles inside probe tip which resulted in unit failed to display good ultrasound image.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
As reported, bent tip, there maybe fluid in the sheating.The issue found during an unspecified procedure.There was no patient harm or injury reported due to the event.No user injury reported.
 
Manufacturer Narrative
Section e3 (occupation) has been updated to reflect the correct information obtained.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the root cause of the damaged probe could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12898712
MDR Text Key284001842
Report Number8010047-2021-15263
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-