Model Number 381823 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported bd insyte¿ autoguard¿ shielded iv catheter had foreign matter.The following information was provided by the initial reporter, translated from (b)(6): "when i opened the cap to puncture and looked at the tip, there was a deposit, so i stopped using it.".
|
|
Manufacturer Narrative
|
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2021-11-24.H6: investigation summary: our quality engineer inspected the sample and photographs submitted for evaluation.Bd received one opened retracted unit for investigation.In addition, five photos were also received for investigation.Upon inspection of the received unit and the photos it was identified that foreign matter is present on the inside of the needle cover.No foreign matter was observed on the needle.The foreign matter was found to be a clear, flexible solid and has a rough surface.Based on the foreign matter being stuck to the inside of the needle cover, the foreign matter can most likely be attributed to the manufacturing or molding process.This type of defect may be caused by multiple factors including environmental factors, incoming material, personnel, and processing (e.G., machine component wear).A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
|
|
Event Description
|
It was reported bd insyte¿ autoguard¿ shielded iv catheter had foreign matter.The following information was provided by the initial reporter, translated from japanese: "when i opened the cap to puncture and looked at the tip, there was a deposit, so i stopped using it.".
|
|