MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 381823
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported bd insyte¿ autoguard¿ shielded iv catheter had foreign matter.The following information was provided by the initial reporter, translated from (b)(6): "when i opened the cap to puncture and looked at the tip, there was a deposit, so i stopped using it.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes, d10: returned to manufacturer on: 2021-11-24.H6: investigation summary: our quality engineer inspected the sample and photographs submitted for evaluation.Bd received one opened retracted unit for investigation.In addition, five photos were also received for investigation.Upon inspection of the received unit and the photos it was identified that foreign matter is present on the inside of the needle cover.No foreign matter was observed on the needle.The foreign matter was found to be a clear, flexible solid and has a rough surface.Based on the foreign matter being stuck to the inside of the needle cover, the foreign matter can most likely be attributed to the manufacturing or molding process.This type of defect may be caused by multiple factors including environmental factors, incoming material, personnel, and processing (e.G., machine component wear).A device history record review showed no non-conformances associated with this issue during the production of this batch.H3 other text : see h10.
 
Event Description
It was reported bd insyte¿ autoguard¿ shielded iv catheter had foreign matter.The following information was provided by the initial reporter, translated from japanese: "when i opened the cap to puncture and looked at the tip, there was a deposit, so i stopped using it.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
katie swenson
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key12898741
MDR Text Key285805477
Report Number1710034-2021-01022
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00382903818235
UDI-Public00382903818235
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K201075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2024
Device Model Number381823
Device Catalogue Number381823
Device Lot Number1053241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2021
Initial Date FDA Received11/30/2021
Supplement Dates Manufacturer Received12/06/2021
Supplement Dates FDA Received12/29/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-