Model Number 284002
Device Problems Device Remains Activated (1525); Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Event Description
It was reported by the sales rep that during a rotator cuff repair surgery on (b)(6) 2021, it was observed that the outflow was not calibrating and the wheel was constantly spinning on the fms vue pump device without using shaver.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event Description
It was reported by the sales rep that during a rotator cuff repair surgery on (b)(6) 2021, it was observed that the outflow was not calibrating and the wheel was constantly spinning on the fms vue pump device without using shaver.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device was received at the service center and evaluated.It was reported that ¿the outflow is not calibrating and the wheel is constantly spinning without using shaver.Team changed tubing and reset shaver but issue continued, keeping the patient's joint bloody¿.Per service reports, no defect was found with the device, therefore this complaint cannot be confirmed.As the reported problem was not confirmed and no defects identified, a root cause for the issue that was experienced by the user cannot be determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key12898948
MDR Text Key285169695
Report Number1221934-2021-03510
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number284002
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/15/2021
Initial Date FDA Received11/30/2021
Supplement Dates Manufacturer Received01/06/2022
Supplement Dates FDA Received01/06/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-