MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AC3 IABP NA/AJLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AC3 IABP NA/AJLA Back to Search Results
Model Number IPN917288
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2021
Type of Reportable Event Malfunction
Event or Problem Description
There was no patient involvement.It was reported that during routine maintenance, an abnormal motor sound was found when the equipment was turned on.A red level 1 alarm displayed on the screen: system error 3, which was detected as stepping motor failure.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
There was no patient involvement.It was reported that during routine maintenance, an abnormal motor sound was found when the equipment was turned on.A red level 1 alarm displayed on the screen: system error 3, which was detected as stepping motor failure.
 
Additional Manufacturer Narrative
Qn#(b)(4).No iabp part or recorder strip was returned to teleflex chelmsford for investigation.The reported complaint of system error 3 is confirmed based on the attached video.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 IABP NA/AJLA
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
kaylia pen
16 elizabeth drive
chelmsford, MA 01824
MDR Report Key12899719
Report Number3010532612-2021-00366
Device Sequence Number5447584
Product Code DSP
UDI-Device Identifier10801902172096
UDI-Public10801902172096
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN917288
Device Catalogue NumberIAP-0601
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2021
Supplement Date Received by Manufacturer12/21/2021
Initial Report FDA Received Date11/30/2021
Supplement Report FDA Received Date12/23/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-