| Model Number |
PNDL, 3BV TVH TP 32G 4MMTC 100CT 30/CS YELLOW |
| Medical Device Problem Code |
Compatibility Problem (2960)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer made several attempts to contact customer to ensure the initial concern has been resolved-unable to establish contact with customer.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for pen needle not aligning with compilable pen needle injector.Customer stated the package had not been open or damaged when received.Customer has been using the product out of this package since (b)(6) 2021.The customer is using compatible product.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the pen needles and no medical intervention related to the use of the product was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 06-jan-2022.H6: updated fda's type of investigation, investigation findings, and investigation conclusions.H10: pen needles were not returned for evaluation.Complaint was forwarded to supplier quality based on complaint's description for investigations.No product was returned to thi, internal evaluation has been completed by the manufacturer, reported defect not reproduced on retention samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
|
| |
|
Search Alerts/Recalls
|