• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; HUMERAL HEAD SHORT, 50MM (BETA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; HUMERAL HEAD SHORT, 50MM (BETA) Back to Search Results
Model Number 314-13-14
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 10/28/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Approximately 8 yrs post an anatomic tsa, this (b)(6) male patient is status post removal of a loose glenoid component and conversion to a hemiarthroplasty was completed on (b)(6) 2020 (need mdr report#).Patient had reported increased pain and failure of bone graft to incorporate in the glenoid.Noted during this implantation - there was a large central contained lesion in the glenoid, glenoid fixation not possible.Defect was bone grafted and implanted back to a stable surface.On (b)(6) 2021, this patient presented with a painful failed hemiarthroplasty.On (b)(6) 2021, this patient was revised and was converted to a hemiarthroplasty and reported the humeral stem was retained.The study reported the surgery date as the ¿resolution¿ date.The case report form indicates this event is possibly related to devices and definitely related to the procedure.This event report was received through clinical data collection activities.
 
Additional Manufacturer Narrative
Pending evaluation.Concomitant device(s): equinoxe replicator plate 4.5mm o/s, cn: 300-10-45, sn: unknown.
 
Event or Problem Description
Approximately 8 yrs post an anatomic tsa, this 56 y/o male patient had an anatomic revision due to glenoid loosening.Defect was bone grafted and implanted back to a stable surface.
 
Additional Manufacturer Narrative
Section h10: (h3) the revision reported was likely the result of both patient conditions and an insufficient bond between the cage glenoid and the bone, which led to aseptic (non-infected) glenoid loosening.However, this cannot be confirmed as the devices and pre-revision x-rays were not available for evaluation.Section h11: *the following sections have corrected information: (b5) describe event or problem: approximately 8 yrs post an anatomic tsa, this 56 y/o male patient had an anatomic revision due to glenoid loosening.Defect was bone grafted and implanted back to a stable surface.(d4) catalog number: 314-13-14, serial number: (b)(6), expiration date: 13-feb-2018, unique identifier (udi) #: (b)(4) (d6a) if implanted, give date: (b)(6) 2013 (d6b) if explanted, give date: (b)(6) 2021 (d11) concomitant device(s): 300-10-15, 2674888 - equinoxe replicator plate 1.5mm o/s 300-20-02, 2604116 - equinox square torque define screw drive kit 310-02-53, 2392928 - equinoxe, humeral head tall, 53mm (beta) (g4) pma/510(k)number: k113309.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
HUMERAL HEAD SHORT, 50MM (BETA)
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key12900381
Report Number1038671-2021-00643
Device Sequence Number14637617
Product Code KWT
UDI-Device Identifier10885862172716
UDI-Public10885862172716
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/13/2018
Device Model Number314-13-14
Device Catalogue Number314-13-14
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer07/05/2022
Initial Report FDA Received Date11/30/2021
Supplement Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age56 YR
Patient SexMale
Patient Weight126 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-