• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number DBP-200SOLID145
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 11/02/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Following treatment of the right superficial femoral artery (sfa) with a diamondback peripheral orbital atherectomy device (oad), the physician retracted the device for removal.Resistance was encountered during removal, and the oad could not be retracted completely.The crown was advanced to relieve the resistance.Fluoroscopy was then used during continued attempts to retract the device.The crown was retracted and became entrapped in the distal portion of the insertion sheath.A technician pulled on the device to bring it back, and the physician pulled on the driveshaft.The oad then came free and was extracted from the patient.However, a 15cm portion of the driveshaft and crown remained in the distal portion of the sheath.The fragment protruded from the hub of the sheath.The physician pulled the fragment out.Angiogram then revealed a type c dissection in the sfa.The dissection was in the area of atherectomy treatment.The procedure was completed and the dissection was resolved with pre-planned angioplasty and stent placement.The patient was stable.
 
Additional Manufacturer Narrative
Device analysis conclusion: the oad was returned to csi for analysis.Visual examination confirmed the reported driveshaft fracture.There was no other damage observed with the driveshaft crown or handle assembly that would have contributed to the reported events.Scanning electron microscopy analysis of the fractured filars shows evidence of fatigue, which is an indication that the driveshaft was spun in a high stress environment.Review of the device data log shows a stall event at the end of the procedure.It is unknown if the stall event occurred during an attempt to remove the device through the sheath or if the stall occurred during the treatment prior to removal.If the device was spun while pulling back through the sheath, this could have led to a stall and an elevated stress condition on the driveshaft, which may have contributed to the fracture.However, this could not be confirmed, and the exact cause of the fracture remains unknown.The report that the device was stuck in the introducer and the report that a dissection occurred could not be confirmed through analysis.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Common Device Name
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
morgan hill
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key12900545
Report Number3004742232-2021-00403
Device Sequence Number14440011
Product Code MCW
UDI-Device Identifier10850000491202
UDI-Public(01)10850000491202(17)230131(10)364356-1
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K190634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Model NumberDBP-200SOLID145
Device Catalogue Number7-10057-04
Device Lot Number364356-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Initial Report FDA Received Date11/30/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient Weight54 KG
Patient EthnicityHispanic
-
-