| Model Number |
314-02-02 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: device evaluated by manufacturer: pending evaluation.Concomitant medical products: stemless humeral head 41mm x 13mm x alpha, cn: 310-62-41, sn: unknown.Stemless humeral comp laser cage, size 1, cn: 300-60-01, sn: unknown.
|
| |
|
Event or Problem Description
|
|
This (b)(6) female patient presented with dissociation of polyethylene liner without traumatic incident.Shoulder is also dislocated anteriorly.And was converted to rtsa of the right shoulder.Outcome reported as resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the revision reported may have been the result of joint instability, which allowed for the glenoid to loosen and disassociate from the glenoid bone and the shoulder to dislocate anteriorly.However, this cannot be confirmed because the components were not returned to exactech for evaluation.
|
| |
|
Search Alerts/Recalls
|