COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Model Number G33016 |
Device Problems
Structural Problem (2506); Device Tipped Over (2589); Insufficient Information (3190)
|
Patient Problems
Pain (1994); Perforation of Vessels (2135); Insufficient Information (4580)
|
Event Type
Injury
|
Event Description
|
Patient allegedly received a gunther tulip filter on (b)(6) 2012 due to trauma, secondary to a motor vehicle accident (mva).
|
|
Manufacturer Narrative
|
The previous mdr was submitted by william cook europe under manufacturer report reference number 3002808486-2021-01506.Additional information provided determined that this device was manufactured by cook inc (cinc).With the submission of this initial report, cinc informs that all future submissions regarding this complaint will be handled under manufacturer report number referenced in this initial medwatch report.Occupation: non-healthcare professional.Investigation: per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.A total of (b)(4) devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Event Description
|
Patient allegedly received the tulip filter via the right common femoral vein.Patient is alleging device tilt and vena cava perforation.Patient notes and further alleges experiencing "pain".
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation, tilt, and pain.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported pain is directly related to the filter and unable to identify a corresponding failure mode at this point in time.A total of 20 devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Additional information: b5, b6 investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.A total of 20 devices were manufactured in the reported lot.To date, no other complaints have been reported against the lot.The associated work order was reviewed.No related/relevant notes were documented.The device is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Per a computed tomography (ct) abdomen: "filter type: cook.Ivc stenosis: no.Filter cone position: below the renal veins.Filter migration: no.Filter fracture/bending: no.Filter tilt: no.Filter penetration: yes.Other findings: no.Impressions: the anterior strut penetrates 8 mm through the ivc wall"."the left strut penetrates 6 mm through the ivc wall".The posterior strut penetrates 8 mm through the ivc wall"."the right strut penetrates 8 mm through the ivc wall".
|
|