|
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the marquee product is bd-santo domingo.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one marquee disposable core biopsy instrument was returned for evaluation.During visual evaluation, the device appeared to have residue and the device was returned in half primed position thus leaving the sample notch in exposed condition.Upon functional testing, the device was able to be fully primed with no issues.The device was further tested by cocking and firing the device 12 times at the 18 mm and 25 mm into a pork shoulder using the automatic and semiautomatic firing options.The device was able to prime and fire properly.All 24 samples were obtained.Therefore, the investigation is unconfirmed for the reported failure to fire issue, as the device was able to fire without issues.However, the investigation is inconclusive for the reported difficult to remove issue, as conditions of use cannot be replicated.A definitive root cause for the alleged failure to fire and difficult to remove issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2024), g3, h6 (method).H11: b5, h6 (device, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|