MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problems
Excess Flow or Over-Infusion (1311); Insufficient Flow or Under Infusion (2182); Infusion or Flow Problem (2964)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider via a company representative regarding a patient receiving morphine 25mg/ml and clonidine via an implantable pump.It was reported that the patient was in the emergency room over this last weekend (11/20 - 11/21) with symptoms common to opioid withdrawal.No alarms are active on the pump.Today (11/23) the patient came into the clinic.The programmer showed an expected reservoir volume of over 5 ml.Actual reservoir volume was zero.There were no found environmental, external or patient factors that may have led or contributed to the issue.The diagnostics and troubleshooting performed was the pump logs were read.A thorough history was taken.The pump reservoir was checked and found to be completely empty.No active alarms or alerts.The actions and interventions taken to resolve the issue was the patient¿s pump was filled today.They were unsure of this was a one time error in programming or partial pocket fill or a case of the pump going bad.The patient¿s symptoms will be closely monitored.His reservoir volume will be checked in 15 days and compared with the expected volume.Surgical intervention did not occur and was unknown if surgical intervention was planned.The issue was not resolved at the time of the report and it was noted that the healthcare provider would not have any further information regarding the event.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 8731sc, serial# (b)(6), implanted: (b)(6) 2009, product type: catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the device manufacturer representative indicated that last week the patient was filled with 20 ml of drug.It was reported that reservoir was checked and the expected volume was 16.1 ml and the actual reservoir volume was only 14 ml.The patient's rate was decreased by 10% and his actual vs expected volumes would be compared again in one week.It was noted that the patient experienced symptoms of overinfusion.They are hoping to schedule this patient for pump replacement on (b)(6) 2021.The plan was to place the pump into minimum rate infusion mode and return it for analysis.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 8731sc lot# serial# (b)(6), implanted: (b)(6) 2009, explanted: product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the device manufacturer representative indicated that during the replacement surgery the c atheter did not aspirate.It was noted that the catheter appeared to be intact and without defect.A section of the pump segment was replaced but it still did not aspirate.Both devices were returned for analysis.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 8731sc, serial# (b)(6), implanted: (b)(6) 2009, product type: catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the device manufacturer representative indicated that the pump was filled with saline and set to minimum rate when it was removed on (b)(6) 2021.It was noted that cultures were taken from the pump pocket and that the devices were returned for analysis.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 8731sc, serial# (b)(6), implanted: (b)(6) 2009, product type catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: the pump and catheter were returned and no significant anomalies were identified.Continuation of d10: product id 8731sc serial# (b)(6) implanted: (b)(6) 2009 product type catheter medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|