• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Model Number IPN048959
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Part number reported, 5-12614 et tube, cuffed, spiral-flex 7.0, is not being manufactured currently, however, another part number from the same family (p/n 5-10315 l/n 73l2100123) was used for the "verification of failure mode reported in the current manufacturing process" and was conducted as follows: 1 (b)(4) samples were taken from the current production; the samples were visually inspected and issue reported "broken/cracked - pilot balloon/valve" was not observed in the current manufacturing process.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported "the doctor found balloon broken when using on patient".No patient injury or harm reported.Patient condition reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
marling heaton
3015 carrington mill blvd
morrisville 27560
MDR Report Key12908287
Report Number3003898360-2021-01053
Device Sequence Number20037657
Product Code BTR
UDI-Device Identifier14026704617071
UDI-Public14026704617071
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K844296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/11/2025
Device Model NumberIPN048959
Device Catalogue Number5-12614
Device Lot Number73B2000277
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2021
Initial Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-