• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; PERCUTANEOUS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OCTRODE LEAD KIT, 30CM LENGTH; PERCUTANEOUS LEAD Back to Search Results
Model Number 3183
Medical Device Problem Codes High impedance (1291); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 03/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Related manufacturer reference number: 3006705815-2021-03148, 1627487-2021-18485.It was reported the patient was experiencing shocking pains at the scs system implantable pulse generator's implant site.Reportedly, the patient fell into a chair and landed on the generator site.Since the patient experiences an intermittent shocking while walking, bending over and doing other physical activities.Surgical intervention took place wherein the system was explanted and replaced to resolve the issue.During the procedure the patient aspirated and the procedure was completed as there was no consequences to the patient.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 30CM LENGTH
Common Device Name
PERCUTANEOUS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12909038
Report Number1627487-2021-18484
Device Sequence Number14673580
Product Code LGW
UDI-Device Identifier05415067017420
UDI-Public05415067017420
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2017
Device Model Number3183
Device Catalogue Number3183
Device Lot Number5262242
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/16/2021
Supplement Date Received by Manufacturer01/29/2022
Initial Report FDA Received Date12/01/2021
Supplement Report FDA Received Date01/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/13/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient Weight79 KG
-
-