• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH WORKING ELEMENT, PASSIVE, FOR RESECTION IN SALINE; WORKING ELEMENTS (MONOPOLAR, BIPOLAR, FOR KNIVES, FOR PROBES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH WORKING ELEMENT, PASSIVE, FOR RESECTION IN SALINE; WORKING ELEMENTS (MONOPOLAR, BIPOLAR, FOR KNIVES, FOR PROBES) Back to Search Results
Model Number WA22367A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 11/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of a bladder tumor (turbt) procedure, the patient was jerking due to electric nerve stimulation.Consequently, spinal anesthesia was used to complete the surgery.No further information was provided, but there was no report of any product malfunction.
 
Additional Manufacturer Narrative
Additional information: d10 - concomitant medical products and therapy dates device evaluation: the suspect medical device was not returned to olympus for evaluation/investigation but to olympus medical systems india (omsi), (returned to omsi on 2021-11-08).The evaluation/investigation did not reveal any malfunction of the working element but found the device to be working correctly.The reported working element is thus rated as meeting its specification.The exact cause of the patient¿s outcome and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, based on the information provided, the reported issue can most likely be attributed to use error.When using a bipolar resectoscope (plasma technology), the current exclusively flows between the two poles of the electrode.Therefore, it is assumed that the nerve stimulation that was triggered in this patient is associated with anatomical conditions and the anesthesia that was chosen.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the working element without showing any abnormalities.The case will be closed on olympus side with no further actions.The reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WORKING ELEMENT, PASSIVE, FOR RESECTION IN SALINE
Common Device Name
WORKING ELEMENTS (MONOPOLAR, BIPOLAR, FOR KNIVES, FOR PROBES)
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662955
MDR Report Key12909252
Report Number9610773-2021-00344
Device Sequence Number3900140
Product Code KNS
UDI-Device Identifier04042761066242
UDI-Public04042761066242
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K030194
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberWA22367A
Device Catalogue NumberWA22367A
Device Lot Number195W
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Supplement Date Received by Manufacturer02/11/2022
Initial Report FDA Received Date12/01/2021
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS ELECTROSURGICAL GENERATOR "ESG-400".
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-