| Catalog Number |
VG0108N |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a hair was found in a vascu-guard product.This was identified before patient use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, e1: initial reporter last name, e1: initial reporter e-mail, d9, h3, h6, h10.B5: it was further reported the hair was found in the jar.H10: the sample was received for evaluation with the tamper-evident label broken.The tissue and a small amount of liquid were present inside the jar.Visual inspection of the sample was performed which identified two particulates ¿ a hair that appeared to be an eyelash and a smaller black fiber present on the tissue.No particulate matter was visible in the liquid remaining in the jar.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|