Model Number 1550250-15
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Angina (1710)
Date of Event 07/09/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Patient id: (b)(6).It was reported that the procedure was performed on (b)(6) 2020, treating lesions in the mid right coronary artery (rca) and in the proximal left anterior descending (lad) artery.A 2.5 x 15 mm xience sierra stent was implanted in the rca and a 3.5 x 38 mm xience sierra stent was implanted in the lad.On (b)(6) 2021, the patient presented to the emergency department with chest pain, relieved with nitroglycerin.A coronary artery bypass graft (cabg) procedure was performed.Post procedure, the patient remained on intra-aortic balloon pump (iabp) for low ejection fraction.The patient was discharged to home on (b)(6) 2021.No additional information was provided.
 
Additional Manufacturer Narrative
The stent remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
The stent remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
Patient id: (b)(6).It was reported that the procedure was performed on (b)(6) 2020, treating lesions in the mid right coronary artery (rca) and in the proximal left anterior descending (lad) artery.A 2.5 x 15 mm xience sierra stent was implanted in the rca and a 3.5 x 38 mm xience sierra stent was implanted in the lad.On (b)(6) 2021, the patient presented to the emergency department with chest pain, relieved with nitroglycerin.A coronary artery bypass graft (cabg) procedure was performed.Post procedure, the patient remained on intra-aortic balloon pump (iabp) for low ejection fraction.The patient was discharged to home on (b)(6) 2021.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of angina is listed in the xience sierra, everolimus eluting coronary stent system (eecss), electronic instructions for use (eifu) as a known patient effect of coronary stenting procedures.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined; however, the reported treatments appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12913287
Report Number2024168-2021-10950
Device Sequence Number12002514
Product Code NIQ
UDI-Device Identifier08717648227097
UDI-Public08717648227097
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/14/2021
Device Model Number1550250-15
Device Catalogue Number1550250-15
Device Lot Number0010841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer12/28/2021
Initial Report FDA Received Date12/01/2021
Supplement Report FDA Received Date01/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Disability;
Patient Age53 YR
Patient SexMale
Patient Weight80 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-