MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV VALVE; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FV410T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 11/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing and quality control data: the progav® was manufactured by a qualified employee in july 2008.Deviations during assembly did not occur.The valve was sterilized by miethke and released for shipment after final inspection.The progav® has a normal pressure range of 0 to 20 cmh2o.The valve was inspected as article fv410t.All parameters have been inspected and approved during the manufacturing process.All parameters have been assessed as meeting specifications.Visual inspection: no significant deformations or damage of the valve were detected during the visual inspection.Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav® 2.0 was tested and is adjustable to all specified pressures.Klick force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of overdrainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve is operating within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, slightly deposits were found in progav®.Result: based on our investigation, we are unable to substantiate the claim of occlusion.At the time of our investigation, the valve was shown to be permeable.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke (b)(4).No further regulatory actions are required from our point of view.
 
Event or Problem Description
It was reported that a progav valve (part # fv410t) was implanted during a procedure performed on (b)(6) 2014.According to the complainant, the shunt system was believed to be operated in overdrainage and difficult to adjust.The patient underwent a revision procedure on (b)(6) 2021.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: (b)(6), height: 156 centimeters (cm), weight: (b)(6), gender: female.It´s the same patient as mdr id # 3004721439-2021-00321.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key12913391
Report Number3004721439-2021-00320
Device Sequence Number14628277
Product Code JXG
UDI-Device Identifier04041906132071
UDI-Public4041906132071
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFV410T
Device Catalogue NumberFV410T
Device Lot Number20008072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2021
Initial Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/30/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age13 YR
Patient SexFemale
Patient Weight48 KG
-
-