ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 45MM 28CM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Model Number PCE45A
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # 217a41.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse45a device was returned with no apparent damage and with no reload present.A dislodge pan was found lodged inside the channel.When the device was returned which leads to the difficulty of the installation of the reload.The pan was removed from the channel and the device was tested for functionality with a test reload.The device fired, cut and formed all the staples as intended.The staple line and cut line were complete and the staples meet the staple release criteria.As part of our quality process, the manufacturing records of this batch was reviewed, and the manufacturing standards were met prior to the release of this batch.It is also possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported that during the segmentectomy of left lung surgery, could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 45MM 28CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key12913686
Report Number3005075853-2021-07307
Device Sequence Number4946283
Product Code GDW
UDI-Device Identifier10705036014461
UDI-Public10705036014461
Combination Product (Y/N)N
PMA/510(K) Number
K110385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2024
Device Model NumberPCE45A
Device Catalogue NumberPCE45A
Device Lot NumberU94M41
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/04/2021
Initial Report FDA Received Date12/01/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-