| Model Number |
OCN005 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
11/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from the healthcare professional (hcp) via the manufacturer representative regarding an event which occurred during an unknown spinal procedure in a patient diagnosed with painful osteoid osteoma.It was reported that while drilling through the thick cortex, hcp tried to reverse drill out of bone because it would not advance, the drill tip sheared off and was left behind stuck in the tibia.A revision surgery will be scheduled to remove the drill tip.No other patient injury / complication was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B2: outcome attributed to adverse event updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Revision surgery performed on (b)(6) 2021 to remove the piece and scraped out the nidus the traditional way to treat the oo instead of using our minimally invasive device procedure details - drilling access with a trocar to access into a young teen tibia for rf treatment of a osteoid osteoma.
|
| |
|
Search Alerts/Recalls
|