| Model Number |
1506-40-004 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical adverse event received for tibial dislocation.Event is serious and is considered severe.Event is probably related to both device and procedure.Date of implantation: (b)(6) 2020, date of event (onset): (b)(6) 20201, (right knee).Treatment: knee revision (b)(6) 2021.The clinical database now reports that there has been a knee revision--right knee was revised for loosening tibial component, right knee--that occurred on (b)(6) 2021.Tibial tray, tibial insert, tibial stem, offset adaptor, and tibial augment were reportedly revised--femur and patella components were retained.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot/ dhr review: 1) quantity manufactured: 12.2) date of manufacture: 26-jun-2019.3) any anomalies or deviations identified in dhr: lot placed on process hold for qa to review clocking lines, lines were present and deemed acceptable.No correlation with the complaint description.4) expiry date: 31-may-2029.5) ifu reference: (b)(4).
|
| |
|
Search Alerts/Recalls
|