It was reported that a patient underwent a prostatectomy procedure on (b)(6) 2021 and suture was used.During the procedure, ng to put the needle-suture through a trocar after the needle was grasped with a needle-holder, the suture was detached from the needle easily.It was not a control release needle.No adverse patient consequences were reported.No additional information was provided.
|
|
Product complaint # (b)(4).Date sent to the fda: 12/29/2021.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary one empty opened foil packet, a paper lid, a winding former with a partially dispensed suture and a needle of product code y305 were returned to ethicon inc for analysis.In order to evaluate the conditions of the returned sample, the swage area and hole of the needle were noted with marks that appears to be by surgical instrument and the hole was observed with suture remnant and flat.To avoid this kind of damage: grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.During the visual inspection of the suture, the end was observed cut by sharp edge.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Functional test was not performed, due to needle was received detached from suture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|