Model Number 9735665
Device Problems Display or Visual Feedback Problem (1184); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/11/2021
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Concomitant medical products: other relevant device(s) are: product id: 9 735764r, lot/serial #: unknown.Product id: 9736356, lot/serial #: unknown.The manufacturer representative went to the site to test the navigation system.The ssd was corrupt.The ssd was replaced.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system.It was reported outside of a procedure that the system froze during navigation.The site stated that both monitors seized up.Troubleshooting was performed and the system did a hard restart but froze again.The next step was to replace the computer.No patient was present at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12916092
MDR Text Key281595713
Report Number1723170-2021-02792
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/11/2021
Initial Date FDA Received12/02/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-