• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES RI ROBOTIC DRILL ATTACHMENT; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10015
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/15/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference case: (b)(4).
 
Event Description
It was reported that during a cori assisted tka surgery the real intelligence robotic drill had several drill time out errors (case: (b)(4)).The tech swapped out robotic drill attachments and continued with the case as planned.After the case, they noticed that an internal component on the ri robotic drill attachment was causing the issues (case: (b)(4)).The procedure was completed, with a non-significant delay, using a s+n back-up device.Patient was not harmed beyond the problem reported.
 
Manufacturer Narrative
H3, h6: the ri robotic drill attachment part number rob10015, serial number (b)(6) , used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file.As a result of the risk assessment the documented risk level has undergone further investigation by the quality team.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with drill attachment wear and tear.Based on the investigation, the need for a corrective action is recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RI ROBOTIC DRILL ATTACHMENT
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12916484
MDR Text Key281596578
Report Number3010266064-2021-00839
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757345
UDI-Public00885556757345
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10015
Device Catalogue NumberROB10015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-