MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 10 DEG E1 LINER 36MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 10 DEG E1 LINER 36MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Off-Label Use (1494); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 11/11/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant products: cat# 110010244 g7 osseoti 3 hole shell 52mm e lot#6964449; cat# 52.34.0688 mathys head (competitor) lot# 2276691; cat# 010000999 g7 screw 6.5mm x 30mm lot#6817834; unknown mathys stem (competitor).Report source: foreign country: (b)(6).The device will not be returned for analysis, as the device was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that the patient underwent another revision surgery approximately 5 months later due to recurrent dislocations.It is noted that competitor products were used with zimmer biomet products during the last revision.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified potentially under sized acetabular cup for the size of the femoral head.In the appropriate clinical context, this could account for recurrence dislocations.Review of the device history records identified no deviations or anomalies during manufacturing.It was identified that zimmer biomet devices were implanted with competitor devices.Zimmer biomet has not confirmed the compatibility for these combinations of devices.It is unknown if these off-label usages may have caused or contributed to the reported events.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10 DEG E1 LINER 36MM E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12916789
Report Number0001825034-2021-03260
Device Sequence Number4994737
Product Code LPH
UDI-Device Identifier00880304526761
UDI-Public(01)00880304526761(17)250303(10)6760350
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/03/2025
Device Model NumberN/A
Device Catalogue Number010000896
Device Lot Number6760350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/11/2021
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date12/02/2021
Supplement Report FDA Received Date03/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight80 KG
-
-