BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3854
Medical Device Problem Codes Break (1069); Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the moderately tortuous and moderately calcified mid circumflex artery.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure, the device would not cross the lesion.A guidezilla guide extension catheter advanced.However, upon inserting the device into the guidezilla, the shaft broke and separated into two pieces while outside of the patient body.The breakage was caused by the kink noted on the shaft.The procedure was completed with another of the same device.There were no complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12917070
Report Number2134265-2021-15169
Device Sequence Number14758052
Product Code NWX
UDI-Device Identifier08714729888260
UDI-Public08714729888260
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/15/2023
Device Model Number3854
Device Catalogue Number3854
Device Lot Number0028027378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/24/2021
Initial Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-