Catalog Number 00234702046
Medical Device Problem Codes Device Damaged by Another Device (2915); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was requested but not available for return.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported during fixation of a plate, the threads of the screw were damaged due to the angle of implantation causing it to run into another screw.The screw was removed and a new one was used to complete the surgery.Attempts have been made and no further information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Visual examination of the provided picture identified that there is damage to the screw threads.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to use error as it was stated that the surgeon deviated from the implantation angle of a screw which caused a collision between two screws.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTICAL BONE SCREW SELF-TAPPING HEX HEAD 3.5 MM DIAMETER 46 MM LENGTH
Common Device Name
PLATE, FIXATION
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12917266
Report Number0002648920-2021-00432
Device Sequence Number19959814
Product Code HRS
UDI-Device Identifier00889024051836
UDI-Public(01)00889024051836(11)121011(10)62166452
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K142579
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00234702046
Device Lot Number62166452
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2021
Supplement Date Received by Manufacturer02/17/2022
Initial Report FDA Received Date12/02/2021
Supplement Report FDA Received Date02/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Patient Age34 YR
Patient SexMale
-
-