SMITH & NEPHEW, INC. UNKN PEEK SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202900
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Damage to Ligament(s) (1952)
Event Date 02/13/2021
Event Type  Injury  
Event Description
It was reported that on literature review "comparison of single versus double-row techniques in arthroscopic rotator cuff repair of full-thickness tears", 4 patients had a re-tear after a double row rotator cuff repair procedure using a footprint pk anchor.This event was resolved by performing a revision surgery.Patient outcome is unknown.No further information is available.
 
Manufacturer Narrative
Internal complaint reference case-(b)(4).Article cite: fatih dogar, erdal uzun.Comparison of single versus double-row techniques in arthroscopic rotator cuff repair of full-thickness tears.Ann clin anal med 2021;12(4):422-427.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.No product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was a isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the complaint revealed there were no patient injuries reported and no apparent patient impact based on the details provided.Therefore, no further medical assessment is warranted.There was no way to determine if the device contributed to the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN PEEK SUTURE ANCHOR
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12917442
MDR Text Key281604092
Report Number1219602-2021-02635
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010652324
UDI-Public03596010652324
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K100159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Literature,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number72202900
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-