Catalog Number 823164 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Hydrocephalus (3272)
|
Event Date 11/17/2021 |
Event Type
Injury
|
Event Description
|
1 of 3 reports.Other mfg report numbers: 3013886523-2021-00535, 3013886523-2021-00536.A physician reported a hakim valve was implanted via v-p shunt on unknown date with unknown setting.The set pressure could be changed to 30 mmh2o, and the flow could be confirmed, but the symptoms of hydrocephalus did not improve, therefore, the valve was removed and replaced on november 17, 2021.The device was used with ns0338(serial;unk) and ns0339 (serial;unk).
|
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Manufacturer Narrative
|
The hakim valve (823164) was returned for evaluation.Failure analysis - the valve was visually inspected; some biological debris was seen in valve casing and a needle hole in the needle chamber.The position of the cam when valve was received was 60mmh2o.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism, at the time of investigation no functional issues were noted.
|
|
Search Alerts/Recalls
|
|