(b)(4).Batch # u5af39.Device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the tx30v device arrived with no apparent damaged and with two reloads present "cartridge bxr30v, t5ev7h was received fully fired and cartridge cxr30v, t5ev7h was received with the drivers partially advanced and with 4 staples protruding, in addition a tyvek was received along with the device.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.It should be noted that product failure is multifactorial.It should be noted that when manipulating the device only the closing trigger should be grasp until ready to fire the device.Do not grasp the firing trigger before the device is to be fired.If the firing trigger is manipulated it could cause the staples to deploy partially or completely resulting in malformed staples and a premature lockout situation.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.
|