• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. QUICK CONNECT GUIDE HANDLE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. QUICK CONNECT GUIDE HANDLE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED Back to Search Results
Catalog Number 110004319
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the guide handle broke during a surgery.According to or nurse, the handle is complete but cannot be used.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Foreign: event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Cosmetic inspection confirms the product has disassembled.The plunger is no longer welded to the body.Cosmetic inspection also found nicks, scratches and worn etching.The fracture site and type is consistent with access quick-connect guide handle plunger fracture complaint summary.Investigation in which overload type of failure was identified.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK CONNECT GUIDE HANDLE
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER, UNCEMENTED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12918848
Report Number0001825034-2021-03243
Device Sequence Number5143279
Product Code MBF
UDI-Device Identifier00887868462979
UDI-Public(01)00887868462979(11)170202(10)900620
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K193038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110004319
Device Lot Number900620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer02/02/2022
Initial Report FDA Received Date12/02/2021
Supplement Report FDA Received Date02/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-