• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C8750
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink system y-type blood/solution set would not disconnect and was observed cracked.It was further reported, when the unspecified infusion was completed ¿the nurse went to disconnect the tubing from the trifusion lumen the tubing was stuck inside the lumen.In attempting to remove it a small piece of plastic from the iv tubing broke off and remained stuck in the lumen." there was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection was performed and the a unspecified defect was noted at the tow piece luer lock assembly.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key12920461
MDR Text Key281757151
Report Number1416980-2021-07096
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8750
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TRIFUSION LUMEN (NON-BAXTER).; UNSPECIFIED SOLUTION
-
-