| Model Number |
19AGN-751 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported a 19mm sjm regent heart valve was chosen for procedure.During the procedure, the user reported while parachuting the valve, it was difficult to seat the valve on the annulus.The device was explanted and exchanged with a smaller, 17mm sjm regent heart valve.The event did lead to prolonged procedure, and prolonged time on the cardiopulmonary bypass.The patient is stable and is recovering.No additional information has been provided.(b)(6) 2021.(b)(6).Oed: hcp was doing avr in the patient and during the procedure hcp implanting 19mm agn valve and while parachuting the valve hcp faced some difficulty to sit the valve on the annulus so he explanted the valve and implant 17mm agn valve.
|
| |
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the valve being difficult to seat on the annulus was reported.The investigation confirmed the device met dimensional and visual specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|