• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1440
Medical Device Problem Codes Misassembled (1398); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2021
Type of Reportable Event Malfunction
Event or Problem Description
The controller dc adapter was returned to the manufacturer because it was unassembled when the box was opened.The controller dc adapter subsequently tested out of specification during manufacturer¿s analysis.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the controller dc adapter ((b)(4)) was returned for evaluation.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned device, as well as visual evidence provided by the site, revealed that the adapter's plug components were present, but were not assembled.This observation did not allow the adapter to provide power.After the components were reassembled, the adapter performed as intended.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed, but not limited, to an improper assembly.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key12920653
Report Number3007042319-2021-08252
Device Sequence Number17460731
Product Code DSQ
UDI-Device Identifier00888707006880
UDI-Public00888707006880
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number1440
Device Catalogue Number1440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/24/2021
Initial Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/27/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
1104 VAD
Patient Age57 YR
Patient SexMale
Patient Weight63 KG
-
-