• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Noise, Audible (3273); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/08/2021
Event Type  malfunction  
Manufacturer Narrative
Device eval by mfr: the jetstream device xc-2.4 was received for analysis.There was approximately 200cc of blood in the waste bag when returned.The shaft and the remainder of the device was inspected for damage.Visual examination showed shaft damage in the form of kink/buckling 1cm from the tip.The functionality of the device was checked by setting up the product per the instructions for use.The device primed as designed.The device was activated, and the blades did not spin as designed.The devices motor was heard; however, no tip rotation was noticed.The devices pod was opened to inspect for damage.It was noticed the pinion gear had slipped off the drive shaft.When the pinion gear slides off the shaft and does not contact the motor gear, rotation of the tip would stop.
 
Event Description
Reportable based on device analysis completed on 19nov2021.It was reported that the device suddenly stopped.A 2.4mm jetstream xc catheter was selected to treat a strongly calcified lesion in the femoral bifurcation.The device was actively used for approximately two to three minutes and then a noise was heard.Following the noise, the device suddenly stopped working and lost all function.The device was removed from the patient and the physician could not reactivate the device.A new jetstream was used to complete the procedure.There were no patient complications.However, device analysis revealed the blades stopped spinning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC ATHERECTOMY CATHETER
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key12920757
MDR Text Key281664867
Report Number2134265-2021-15034
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/02/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0027755155
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-