BOSTON SCIENTIFIC CORPORATION JETSTREAM XC ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY
|
Back to Search Results |
|
Model Number 45007 |
Device Problems
Noise, Audible (3273); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device eval by mfr: the jetstream device xc-2.4 was received for analysis.There was approximately 200cc of blood in the waste bag when returned.The shaft and the remainder of the device was inspected for damage.Visual examination showed shaft damage in the form of kink/buckling 1cm from the tip.The functionality of the device was checked by setting up the product per the instructions for use.The device primed as designed.The device was activated, and the blades did not spin as designed.The devices motor was heard; however, no tip rotation was noticed.The devices pod was opened to inspect for damage.It was noticed the pinion gear had slipped off the drive shaft.When the pinion gear slides off the shaft and does not contact the motor gear, rotation of the tip would stop.
|
|
Event Description
|
Reportable based on device analysis completed on 19nov2021.It was reported that the device suddenly stopped.A 2.4mm jetstream xc catheter was selected to treat a strongly calcified lesion in the femoral bifurcation.The device was actively used for approximately two to three minutes and then a noise was heard.Following the noise, the device suddenly stopped working and lost all function.The device was removed from the patient and the physician could not reactivate the device.A new jetstream was used to complete the procedure.There were no patient complications.However, device analysis revealed the blades stopped spinning.
|
|
Search Alerts/Recalls
|
|
|