| Model Number |
ST266T |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to aesculap ag that a s4 element polyaxial screw 6.0x55mm (part # st266t) was used during a procedure performed on an unknown date.According to the complainant, the device disintegrated during surgery.During the operation, surgeon wanted to turn the head of the screw so that it was inserted into the patient and the screw "disintegrated", the inner sealing ring was loosened.The patient was not damaged, the surgeon had to remove the screw and use a new one.The complaint device was returned to the manufacturer for evaluation.An additional medical intervention was necessary.Additional information was not provided nor available.The adverse event is filed under aag reference (b)(4).Associated medwatch reports: 9610612-2021-00756 - 400526775 and 9610612-2021-00758 - 400527068.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results/ visual investigation: a visual examination of the pedicle screws ("a" and "b") and their inlays was performed.No deviation to the specification was identified.The catches for both screws are in the correct position.The inner side of the screw heads showed no wear or signs of usage.Batch history review: there was no deviation to the specification.Conclusion and measures / preventive measures: the received screw shows no deviation to its specification.The complaint description could not be confirmed.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|