This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation and device evaluation.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the tissue pad was peeled and torn away.In addition, the coating of the grasping section was partially peeled off and the coating of the probe was partially peeling.There was a mark in the probe which came in contact with the non-insulated area of grasping section and was abraded against it.However, these defects are not considered severe enough to cause a potential adverse event.The probe check was successful.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the falling of the tissue pad: 1.Due to performing output while grasping nothing (including the case the user kept activating after cutting tissue), the distal end of the tissue pad was peeled and torn away.2.The tissue pad was fell off due to load applied to the tearing tissue pad.The coating of the grasping section could have come off when dirt such as burn was scraped off with something hard.This following information is included in the instructions for use (ifu): "do not activate output in seal and cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating." "when cutting and vessel sealing is performed in seal and cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." ¿if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ olympus will continue to monitor the performance of this device.
|